Description
Condition: For parts or not working
Condition Description: Please Read**** This unit was purchased and returned for the following reasons. It was calibrated tested and calibrated by Keysight on 11/27/23 but there are some reported issues from the last group that purchased this unit. See below. Comes as seen in the images with no additional accessories. 1- Touch Screen does not work, 2- 82357 Server Utility Error, 3- Phase Noise Measurement Variance – During testing, we observed phase noise measurements that deviated from expected benchmarks:- At 10 MHz, phase noise floor values beyond 100 kHz offset were approximately 2–3 dB worse than those measured with our reference unit (Agilent E5052B).- At 1 GHz, floor values were approximately 5 dB worse than expected, based on comparative testing against our in-house analyzer.
MPN: E5052B
Custom Label: 0-554B-78874-MG-CH